Nikon A1 — NIS-Elements MVP

Minimum Viable Protocol for the Nikon A1 confocal on the shared Ti2 body in Flux 4.102 — the day-to-day NIS-Elements workflow once your A1 introduction is done. For the power-up sequence (which is shared with the Abbelight SAFe-360 on the same body), see the Abbelight SOP.

Companion SOP: ICMS-SOP-AMF-001. Software: NIS-Elements AR 5.21.03.

1. Introduction

The Nikon A1 is Nikon’s older generation of confocal laser scanning microscope, retained at the AMF as the everyday confocal alongside the Abbelight super-resolution platform on the same Ti2 body. Excitation is handled by the LU4 laser combiner, fiber-coupled into the scan head; up to four fluorophores can be acquired in the same sample.

EquipmentAbbelight / Nikon A1 microscopy system, Flux 4.102
SoftwareNIS-Elements AR 5.21.03
Companion SOPICMS-SOP-AMF-001
Important. The Nikon A1 PC is not connected to the internet and has no login password. Use NIS-Elements only for acquisition. Local storage is on D, E, F drives. Transfer data via a virus-checked external USB drive.

Lasers (LU4 combiner): 405, 488, 561, 640 nm primary excitation channels; the LU4A Pad in NIS-Elements drives the discrete laser lines. Objectives (shared with the Abbelight platform): 60× TIRF and 100× TIRF. The system has an Okolab Top-stage Incubator for live-cell imaging (temperature, humidity, CO2); see the companion SOP for incubator setup.

2. Practice guide: step by step

After completing the power-up sequence (see the Abbelight SOP, section 2b for the Nikon A1 path), unlock the screen with Ctrl+Alt+Delete then Enter (no password). Open NIS-Elements AR 5.21.03 from the desktop. The system calibrates audibly during initialisation.

Strategy. Find and focus on a robust feature first. If your sample bleaches quickly, configure all acquisition settings (steps 2–4) before locating the region of interest.

Step 1. Find the sample and focus (brightfield)

  1. Ensure all lasers are off. Open the chamber lid and locate the sample through the eyepiece using the side knob.
  2. Once focused, press PFS (Perfect Focus System) to lock focus while you navigate with the joystick.
  3. Close the chamber lid. Switch from ocular view to data acquisition by pressing the Camera L button on the microscope body. It should turn green, confirming the PC-camera link.
Tip. To change sample or well: press ESC, swap the sample, disable ESC, then re-engage PFS.

Step 2. Configure acquisition channels

  1. Set the acquisition mode (simultaneous or sequential channel acquisition).
  2. In the A1 Plus Compact GUI panel, click the colour-pad button to open the Dye and Spectral Setting window.
  3. For each channel, choose the dye and emission/excitation wavelength, select the detector type (standard: DU4), and verify each spectral window matches the expected emission spectrum without overlap.
Compact GUI with channel config: DAPI 405, 488, DsRed2 561, Alx647 640
Channel configuration in the A1 Plus Compact GUI (DAPI 405, 488, DsRed2 561, Alx647 640).

Step 3. Set scanning parameters

LU4A Pad configuration

  1. The beam switcher is set to Confocal (right button under “Beam Switcher”).
  2. All laser-related shutters are open (yellow circles in the interface).
  3. Remove the interlock when ready to scan. When the interlock indicator is red, the safety system is active and lasers are disabled when the chamber lid is open.
LU4A Pad with laser sliders, beam switcher, shutters, interlock
LU4A Pad — laser power sliders, beam switcher, shutters and interlock.

Scanning

ParameterDescription
Scanner SpeedFrames/sec (full images per second). Pixel dwell time = how long the laser sits on each pixel.
SizeImage size in pixels. With high-NA lenses, use 1024×1024 or larger to avoid undersampling.
PinholeTrade-off between optical sectioning and signal strength. Application-dependent.
HV (gain)PMT voltage. Increase to compensate for low laser power and boost detector sensitivity.
OffsetAdjust to suppress background.
  1. Start with low laser power (avoid photobleaching) and moderate HV.
  2. Click Scan to begin live scanning. Look for a well-utilised dynamic range: visible signal, no saturated pixels.
  3. If signal is weak, increase laser power incrementally; monitor for photobleaching.
A1 Plus Compact GUI with scanning and laser parameters
A1 Plus Compact GUI — scanning controls, channel sliders and laser power.

Step 4. Optimise Z-stack for 3D imaging

  1. Scroll along Z manually to identify where the signal begins and ends.
  2. Click Top and Bottom in the Z-stack panel to set the range. Add ±1–2 µm margin.
  3. Set the step size. NIS suggests an optimal value; you can override. Press Run Now to start 3D acquisition.

Final acquisition and saving

Review all parameters before final acquisition. Save in .nd2 format — it preserves all acquisition metadata.

3. Common errors checklist

Error / symptomLikely cause
Saturation (white pixels)Gain or laser power too high — loss of information in bright regions.
Noisy, grainy imageHV (gain) too high relative to signal.
Sample fading over timePhotobleaching from excessive laser power or repeated scanning.
Loss of optical sectioningPinhole set too open.
Image too darkPinhole too closed with weak signal — insufficient light reaching the detector.
Poor resolution despite 100× objectiveImage format too small (e.g. 512×512), causing undersampling.