Minimum Viable Protocol for the Nikon A1 confocal on the shared Ti2 body in Flux 4.102 — the day-to-day NIS-Elements workflow once your A1 introduction is done. For the power-up sequence (which is shared with the Abbelight SAFe-360 on the same body), see the Abbelight SOP.
Companion SOP: ICMS-SOP-AMF-001. Software: NIS-Elements AR 5.21.03.
1. Introduction
The Nikon A1 is Nikon’s older generation of confocal laser scanning microscope, retained at the AMF as the everyday confocal alongside the Abbelight super-resolution platform on the same Ti2 body. Excitation is handled by the LU4 laser combiner, fiber-coupled into the scan head; up to four fluorophores can be acquired in the same sample.
| Equipment | Abbelight / Nikon A1 microscopy system, Flux 4.102 |
|---|---|
| Software | NIS-Elements AR 5.21.03 |
| Companion SOP | ICMS-SOP-AMF-001 |
Lasers (LU4 combiner): 405, 488, 561, 640 nm primary excitation channels; the LU4A Pad in NIS-Elements drives the discrete laser lines. Objectives (shared with the Abbelight platform): 60× TIRF and 100× TIRF. The system has an Okolab Top-stage Incubator for live-cell imaging (temperature, humidity, CO2); see the companion SOP for incubator setup.
2. Practice guide: step by step
After completing the power-up sequence (see the Abbelight SOP, section 2b for the Nikon A1 path), unlock the screen with Ctrl+Alt+Delete then Enter (no password). Open NIS-Elements AR 5.21.03 from the desktop. The system calibrates audibly during initialisation.
Step 1. Find the sample and focus (brightfield)
- Ensure all lasers are off. Open the chamber lid and locate the sample through the eyepiece using the side knob.
- Once focused, press PFS (Perfect Focus System) to lock focus while you navigate with the joystick.
- Close the chamber lid. Switch from ocular view to data acquisition by pressing the Camera L button on the microscope body. It should turn green, confirming the PC-camera link.
Step 2. Configure acquisition channels
- Set the acquisition mode (simultaneous or sequential channel acquisition).
- In the A1 Plus Compact GUI panel, click the colour-pad button to open the Dye and Spectral Setting window.
- For each channel, choose the dye and emission/excitation wavelength, select the detector type (standard: DU4), and verify each spectral window matches the expected emission spectrum without overlap.
Step 3. Set scanning parameters
LU4A Pad configuration
- The beam switcher is set to Confocal (right button under “Beam Switcher”).
- All laser-related shutters are open (yellow circles in the interface).
- Remove the interlock when ready to scan. When the interlock indicator is red, the safety system is active and lasers are disabled when the chamber lid is open.
Scanning
| Parameter | Description |
|---|---|
| Scanner Speed | Frames/sec (full images per second). Pixel dwell time = how long the laser sits on each pixel. |
| Size | Image size in pixels. With high-NA lenses, use 1024×1024 or larger to avoid undersampling. |
| Pinhole | Trade-off between optical sectioning and signal strength. Application-dependent. |
| HV (gain) | PMT voltage. Increase to compensate for low laser power and boost detector sensitivity. |
| Offset | Adjust to suppress background. |
- Start with low laser power (avoid photobleaching) and moderate HV.
- Click Scan to begin live scanning. Look for a well-utilised dynamic range: visible signal, no saturated pixels.
- If signal is weak, increase laser power incrementally; monitor for photobleaching.
Step 4. Optimise Z-stack for 3D imaging
- Scroll along Z manually to identify where the signal begins and ends.
- Click Top and Bottom in the Z-stack panel to set the range. Add ±1–2 µm margin.
- Set the step size. NIS suggests an optimal value; you can override. Press Run Now to start 3D acquisition.
Final acquisition and saving
Review all parameters before final acquisition. Save in .nd2 format — it preserves all acquisition metadata.
3. Common errors checklist
| Error / symptom | Likely cause |
|---|---|
| Saturation (white pixels) | Gain or laser power too high — loss of information in bright regions. |
| Noisy, grainy image | HV (gain) too high relative to signal. |
| Sample fading over time | Photobleaching from excessive laser power or repeated scanning. |
| Loss of optical sectioning | Pinhole set too open. |
| Image too dark | Pinhole too closed with weak signal — insufficient light reaching the detector. |
| Poor resolution despite 100× objective | Image format too small (e.g. 512×512), causing undersampling. |